Standard

NS-EN ISO 32543-3:2026

Published

Corrigenda and amendments are bought separately.

Language
Services

Abstract

This document specifies a method for the measurement of focal spot dimensions within the range of 5 µm to 300 µm of X-ray systems up to and including 225 kV tube voltage. This determination is based on the evaluation of an image with a dedicated focal spot that has been radiographically recorded using an edge and evaluated with a digital method.The imaging quality and the resolution of X-ray images depend highly on the characteristics of the effective focal spot, in particular the size and the two-dimensional intensity distribution as seen from the detector plane.For the characterisation of commercial X-ray tube types (i.e. for advertising or trade) the specific maximum values of annex A should be preferred.

Document information

  • Standard from SN
  • Published:
  • Edition: 2026-01
  • Version: 1
  • Document type: NAT
  • ICS 19.100

Product Relations