Abstract

This European Standard specifies a method for the measurement of focal spot dimensions within the range of 5 um to 300 um of X-ray systems up to and including 225 kV tube voltage, by means of radiographs of sharp edges. The image quality and the resolution of X-ray images highly depend on the characteristics of the focal spot. The imaging qualities of the focal spot are based on the two dimensional intensity distribution in the object plane. For certification purposes the radiographic technique is used.